• 2-channel-instrument, for Pt100 according EN 60751, • Simultaneous display of two measured values or differential measurement • Recording maximum, minimum, hold and average values(average over user selectable time) • Integrated calibration function(According EN60751, up to 20 points) • High measuring accuracy (+ 0,010 °C) • Resolution of 0.001°C below 200°C • Password protection for calibration function • USB interface • Manufacturers test certificate included • Data Logger function for up to 4000 measurements • Optional DKD or ISO- system calibration certification.

Inputs

Pt100 according to EN 60751

Measuring range

 

Pt100

-200°C...+850°C

Accuracy

 

Pt100

±0,015°C from -50°C...+199,99°C
  ±0,025% of reading for the remaing range

Resolution

0,001°C from -200°C...+200°C
  remaining range 0,01°C

EX-mark

-

Memory

4000 measurements

Measuring interval

Adjustable: 1 second...30 minutes

Output

USB

Connectors

8-pole DIN 45326

Working temperature

0°C...+40°C

Display

2-line LCD

Housing

plastic (ABS)

Dimensions

200 x 85 x 40 mm (LxWxH)

Weight

300 g

Power supply

9 V battery

Batterie life

appr. 20 h